·Testing sample varieties: soft magnetic ferrite, permalloy, amorphous, amorphous and nm crystal.
·Test sample shapes: annular, E and U shapes.
·Test samples with closed magnetic circuit can be directly winded and measured on samples. Sample, magnetizing coil (N1) and measuring coil (N2) form a no-load transformer.
·Non-inductive resistance connected on magnetizing coil loop to determine magnetizing current and magnetic field intensity through the measurement of pressure drop on non inductive resistance.
·Magnetic induction obtained through digital integration of coil measuring voltage, magnetic induction peak value locked through digital feedback, magnetic induction lock precision 0.5%.
·Power source and sampling amplifier integrated into a casing with simple interface: one RS232 interface connected to the computer, two way voltage alarm connected to high speed A/D card.
·Different power sources can be selected according to different test requirements: 1kHz ~ 300kHz or 500kHz, 5kHz or 10kHz ~ 100kHz, or other frequency ranges customized.
·Measuring dynamic hysteresis loop through volammetry and digital integration can accurately measure dynamic magnetic characteristic parameters such as μa, δ, Pc, Br and Hc, and calculate μ', μ'', μL, μR, Q and AL, etc.
·Automatic and continuous measurement up to 255 testing points, the testing time of every testing point is about 12 seconds, fixed-frequency or fixed Bm optional in multipoint test.
|